background image
OPTICAL BACKSCATTER REFLECTOMETERTM
(Model OBR 4600)
Extended Range Spot Scan Interrogation1
www.lunainc.com
3157 State Street | Blacksburg, VA 24060
solutions@lunainc.com
Support phone: 1.866.586.2682
1
Extended range sensing works best with a low net change in accumulated strain between the unit and the sensing region.
2
Internal wavelength calibration using a NIST-traceable HCN gas cell occurs upon instrument configuration, but may drift with
temperature. Wavelength offset measured at 1550nm. Thermal drift,
ΔT, in degrees Celcius.
3
Based on maximum wavelength range of 3.2 nm.
4
Actual temperature range will be limited by the sensing fiber used.
5
Temperature accuracies are calculated from the wavelength accuracy using the spectral shift of Rayleigh scatter, 1 GHz = 0.8 °C
= 6.
58 µε. [Othonos and K. Kalli, Fiber Bragg Gratings (Artech House, Boston, 1999)].
6
Larger gage lengths may be required to meet noise specifications in the presence of vibration.
PARAMETER
SPECIFICATION
UNITS
Maximum Range
2,000
meters
Spot Scan Size
80
meters
Maximum Wavelength Range
3.2
nm
Time Per Scan
6.5
seconds
2-pt Spatial Resolution
0.25
mm
Wavelength Accuracy2
± 5.0
+ 12*ΔT
pm
Strain Range3
± 1,250
µStrain
Strain Accuracy2,5
± 4.1 + 9.9
*ΔT
µStrain
Temperature Range3,4
± 175
°C
Temperature Accuracy2,5
± 0.5 + 1.2
*ΔT
°C
Minimum Gage Length6
3
cm
Minimum Sensor Spacing
0.6
mm
The
Extended
Range
Interrogation
capability is available on the OBR 4600
with purchase of the extended range
mode and spot scan mode. To perform
strain and temperature measurements,
the distributed sensing option is required
in addition to the above. This capability
extends
strain
and
temperature
measurements up to a range of 2 km for
applications located in a benign vibration
environment.
APPLICATIONS
•
Evaluating spools of Fiber Bragg
Gratings
•
Evaluating optical networks and
devices
•
Strain and temperature at a distance



Html Pages

1 2 3 4


Datasheet Descarga

Go To PDF Page


Enlace URL



¿ALLDATASHEET es útil para Ud.?  [ DONATE ] 

Todo acerca de Alldatasheet   |   Publicidad   |   Contáctenos   |   Política de Privacidad   |   Enlace a la hoja de datos    |   Intercambio de Enlaces   |   Lista de Fabricantes
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com